Page 38
24- Using the bulb, pump until the pressure on the manometer is slightly above 30 mmHg. Release
the pressure slightly using the release knob on the bulb. When the pressure drops back to 30
mmHg, click Continue.
Note: Once pumping starts ensure to click Continue within 3 minutes. If not, a message will appear
informing that the calibration was unsuccessful. Then close the SphygmoCor XCEL Calibration Software,
turn off/on the module and repeat the calibration procedures from Step 1. Please refer to Restarting the
Calibration Procedure Section in the Calibration Troubleshooting section.