TECH 196 / 196 M
14
en
Calibration
1. Switch on the electronic spirit level. Use the vial to
accurately align the unit to a wall, for example, until
the vial bubble is in the middle between the vial rings.
If mainly used for vertical measurements, the calibration
can also be undertaken with the vertical vial.
2. Hold the electronic spirit level in this position and press
the CAL button.
The CAL display shows calibration mode.
3. Calibration starts when the CAL button is pressed again.
Newly calibrated value with
deviation of ≥ 1° from
factory setting
Recalibrate spirit level
Vibration during
calibration
Recalibrate spirit level
Calibration completed
successfully
Spirit level ready
The reverse test checks the calibration.
Angle ≤ 0.1° to normal
position
Spirit level ready
18
5 sec