7 — Troubleshooting
128
10057344_H
AMSCO® 5000 Series
Operator Manual
7.26 Alarm# 24 Load Door Did Not Open
Condition
Non-critical. Alarm is triggered when load door is detected as closed after automatic door opening sequence (eight
seconds).
General Troubleshooting
Additional Information
Shut-off valve is closed.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Verify if air supply shut-off valve is open.
4. If situation reoccurs, contact STERIS.
Door gutter is obstructed.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Verify there is no obstruction in door gutter.
4. If situation reoccurs, contact STERIS.
7.27 Alarm# 25 Load Door Did Not Close
Condition
Non-critical. Alarm is triggered when load door is not detected as closed after automatic door closing sequence (twelve
seconds).
General Troubleshooting Additional Information
Shut-off valve is closed.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Verify is air supply line shutoff valve is open.
4. If situation reoccurs, contact STERIS.
Door path is obstructed.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Verify there is no object obstructing door path and
try to close door again.
4. If situation reoccurs, contact STERIS.