7 — Troubleshooting
130
10057344_H
AMSCO® 5000 Series
Operator Manual
7.30 Alarm# 28 Unload Door Did Not Close
Condition
Non-critical. Alarm is triggered when unload door is not detected as closed after automatic door closing sequence
(twelve seconds).
General Troubleshooting
Additional Information
Shut-off valve is closed.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Verify is air supply line shutoff valve is open.
4. If situation reoccurs, contact STERIS.
Door path is obstructed.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Verify there is no object obstructing door path and
try to close door again.
4. If situation reoccurs, contact STERIS.
7.31 Alarm# 29 Unload Door Obstruction
Condition
Non-critical. Alarm is triggered when unload door obstruction bar detected an obstruction during automatic door
closing.
General Troubleshooting
Additional Information
Door obstructed.
1. Press Silence Buzzer on touch screen to stop alarm
buzzer.
2. Press Ack on touch screen to acknowledge alarm.
3. Clear door path from obstruction and try to close
door again.
4. If situation reoccurs, contact STERIS.