4 Operating Instructions
54
10057344_H
AMSCO® 5000 Series
Operator Manual
4.11 Aborting Cycle Operation
WARNING
BURN HAZARD
In case of an emergency, always stop cycle by pressing Emergency Stop pushbutton
and wait for water flow to stop prior to trying to open door. Wear appropriate Personal
Protective Equipment (PPE) whenever reaching into the chamber.
If cycle needs to be aborted for any reason, proceed as follows:
1. Press Abort on touch screen. Display shows:
2. Press YES on touch screen to abort cycle. Washer/disinfector performs Abort sequence.
3. Washer/ disinfector drains and display returns to Ready screen.
IMPORTANT: If cycle is aborted before the Thermal Rinse phase is completed, load access code might be required to
open load door. Load access code can be enabled or disabled in the Miscellaneous menu of the Supervisor mode (refer
to Section 5.3, Supervisor Mode Menus).
4.12 Shutdown
WARNING
ELECTRIC SHOCK AND/OR BURN HAZARD
Placing washer/disinfector Power (ON/OFF) toggle switch in OFF position to turn off
washer/disinfector DOES NOT cut off electrical power.
Washer/Disinfector should remain ON at all times except when performing maintenance or repairs, then operator
needs to:
Press bottom portion of Power (ON/OFF) toggle switch (located on load side of washer/disinfector) to turn OFF control
system.
Lockout/ tagout main electrical disconnect switch to OFF position.