7 — Troubleshooting
144
10057344_H
AMSCO® 5000 Series
Operator Manual
7.60 Sump Safety Level Test Required
Condition
Reminder is displayed when number of cycles set for sump safety level test (1000) has been reached. Reminder is
displayed again after five more cycles completed.
General Troubleshooting
Additional Information
Safety level test required.
1. Press Ack on touch screen to acknowledge
reminder.
2. If you want to perform test, answer YES and
control automatically goes into sump safety level
test menu.
3. If you do not want to perform test, answer NO and
control automatically returns to Ready screen.
7.61 Touch screen does not work properly
Condition
Touch screen does not work properly.
General Troubleshooting
Additional Information
Faulty touch screen electrical connection.
1. Restart washer/disinfector using Power (ON/OFF)
toggle switch.
2. Contact STERIS.
Touch screen lost calibration.
1. Restart washer/disinfector using Power (ON/OFF)
toggle switch.
2. Contact STERIS.
Touch screen is damaged.
1. Restart washer/disinfector using Power (ON/OFF)
toggle switch.
2. Contact STERIS.
7.62 Transfer does not occur between conveyors and chamber
Condition
If washer/disinfector is equipped with a SCS conveyor system, transfer does not occur between conveyors and
chamber.
General Troubleshooting
Additional Information
Communication cable is not connected between
washer/disinfector and conveyors.
1. Contact STERIS.
Faulty communication between washer/disinfector and
conveyors.
1. Contact STERIS.