10 — Appendix
24
Operator Manual LED202 In-Ceiling Lighting System
11042437_A
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The STERIS Corporation LED202 SYSTEM, models LED202X2 and LED202X1 is intended for use in the electro-
magnetic environment specified below. The customer or the user of the STERIS Corporation LED202 SYSTEM,
models LED202X2 and LED202X1 should assure that it is used in such an environment.
Immunity test
IEC 60601 4th
Edition Test Level
Compliance Level
Electromagnetic Environment –
Guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst IEC
61000-4-4
±2 kV for power
supply lines
±2 kV for power
supply lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to line
(s)
±2 kV line(s) to earth
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply in-put
lines IEC 61000-4-
11
0% U
T
for 0.5 cycle at
0°, 45°, 90°, 135°,
180°, 225°, 270°
and 315°
0% U
T
; 1 cycle and
70% UT, 25/30
cucles Single phase:
at 0°
0% U
T
; 250/300 cycle
0% U
T
for 0.5 cycle at
0°, 45°, 90°, 135°,
180°, 225°, 270°
and 315°
0% U
T
; 1 cycle and
70% UT, 25/30
cucles Single phase:
at 0°
0% U
T
; 250/300
cycle
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the STERIS
Corporation LED202 SYSTEM, models
LED202X2 and LED202X1 requires
continued operation during power mains
interruptions, it is recommended that
the STERIS Corporation LED202
SYSTEM, models LED202X2 and
LED202X1 be powered from an
uninterruptible power supply or a
battery.
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
30 A/m 30 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE: U
T
is the AC mains voltage prior to application of the test level.