T6800 Rev B Operator Manual Quality Assurance
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
8-8
HOW TO CHECK THE DIAGNOSTIC CYCLE STATUS
NOTE: The STATUS touch pad is operable only when the SYSTEM 1 endo
Processor is in-cycle.
To check the status of the cycle while the “OPERATING” screen is displayed
(while the unit is in-cycle), press the STATUS touch pad. The display will show
one of the following screens, depending on what part of the phase the cycle is
in when the STATUS touch pad is pressed:
STEP 4
... and the clock timer will continue to count down time remaining in the
DIAGNOSTIC cycle.
To return to the “OPERATING” screen, press the STATUS touch pad again, or
wait for the display to return on its own when it automatically times out in 20
seconds.
INFLATE SEAL
TEST
INLET TEMP
TEST
FILL TIME
TEST
PRESS TRANSDUCER
TEST
CIRC PRESSURE
TEST
TC1 & TC2
TEST
CIRC VALVE
TEST
HEAT RATE
TEST
DRAIN VALVE
TEST
DRAIN TIME
TEST
MAXLIFE FILTER
TEST