3. Remove the filter bag from the tank and turn it upside down over a waste container.
4. Empty the debris into the container and clean the filter bag.
5. Replace the filter bag.
Note
Dispose of swarf according to the current local safety regulations for handling and
disposal of swarf and cooling fluid additive.
7.2.2 Filter tube
Check and, if necessary, replace the filter tube.
Note
Dispose of swarf according to the current local safety regulations for handling and
disposal of swarf and cooling fluid additive.
Hint
The combination of the metallic swarf (cutting debris) from metals with a large
difference in electro-positivity can result in exothermic reactions when favorable
conditions are present. This depends on whether a large amount of swarf is produced
during cutting/grinding on the same machine, and on the metals being cut
Examples:
The following are examples of such metal combinations:
aluminum and copper
zinc and copper
7.2.3 Magnetic filter
Check and, if necessary, clean the magnetic filter.
1. Remove the filter from its connection.
2. Slide the plastic outer tube off the magnet.
3. Use a stiff brush to clean the plastic tube.
4. Remount the plastic tube.
5. Remount the magnetic filter back on its connection.
7.2.4 Static filter
Check and, if necessary, clean the static filter.
1. Remove the pump.
2. Loosen the wing nut and remove the static filter.
3. Use a stiff brush to clean the mesh.
4. Rinse the static filter.
5. Remount the static filter.
24 Cooli System
7 Maintenance and service

Related product manuals