Subject Specifications
Power supply Voltage 100-240V/50/60Hz
Power inlet 1-phase (N+L1+PE)
The 2-pin plug (European Schuko) or 3-pin plug
(North American NEMA) is for use on single-
phase electrical power connections.
The electrical installation must comply with
“Installation Category II”.
Power consumption 710 W
Power, idle 5 W
Current, nominal
load
8.7 A
Current, maximum
load
17.4 A
Residual Current
Circuit Breaker
(RCCB)
Type AC 30 mA (or better) is recommended
Operating
environment
Surrounding
temperature
4-40 °C (41-104 °F)
Humidity 10-85 % RH non-condensing (10-85 % RH non-
condensing)
Storage Surrounding
temperature
0-60 °C (32-140 °F)
Humidity 10-85 % RH non-condensing (10-85 % RH non-
condensing)
9.3 Safety Circuit Categories/Performance Level
Safety Circuit
Categories/Performance
Level
Emergency stop
from main machine
PLc, Category 1
Stop category0
9.4 Noise and vibration levels
Noise level A-weighted sound
emission pressure
level at workstations
L
pA
= 71.1 dB(A) (measured value)
(In combination with: Labotom-5 cutting an
aluminum alloy rod Ø 40 mm.)
Uncertainty K = 4 dB
Measurements made in accordance with EN
ISO 11202
30 Cooli System