Troubleshooting Tips
For a complete list of troubleshooting tips, please
consult the LIFEPAK 15 Operating Instructions
LIFEPAK 15 Monitor/Debrillator
Pacing
Observation Action
Failure to capture Increase pacing current
Consider moving pacing electrode
to another location
Oversensing Change to a different ECG lead or
reposition the ECG electrode
Noisy ECG signal Conrm adequate skin preparation
Move the ECG electrodes farther
away from the pacing electrodes
Consider selecting another ECG lead
Monitor screen
displays distortion
while pacing
Reposition electrodes away from
pacing electrodes or select another
lead (I, II or III)
Intrinsic QRS
complexes not sensed
when pacing
Increase ECG size or select another lead
Adjust PPM
18 : 20 01 MAY 95 PACE @ 80, 45 MA LEAD I I X 1 . 0