Page 22
Troubleshooting Tips
For a complete list of troubleshooting tips, please
consult the LIFEPAK 15 Operating Instructions
LIFEPAK 15 Monitor/Defibrillator
NIBP
Observation/
Messages
Action
NIBP TIME OUT • Apply appropriately sized cuff and
properly align cuff artery markings
to extremity
• Repeat measurement
NIBP WEAK PULSE Check for pulse distal to cuff and ensure
cuff ts snugly on the patient’s arm
NIBP CHECK CUFF • Apply appropriately sized cuff and
properly align cuff artery markings
to extremity
• Repeat measurement
NIBP MOTION Have patient lie quietly with extremity
relaxed and supported
NIBP AIR LEAK • Apply appropriately sized cuff and
properly align cuff artery markings
to extremity
• Check that cuff/monitor connection
is secure
• Check cuff for leaks