5. Highlight PPrreessss ttoo sseett and press EEnntteerr to set the product clock.
AAccttuuaattoorr mmaaiinntteennaannccee
CCAAUUTTIIOONN -- Always make sure that the product is clear of obstacles before you use motion functions.
Actuator maintenance allows you to bypass the set limits of the product actuators for testing and troubleshooting.
To access actuator maintenance:
1. Enter the
Maintenance menu
(page 20).
2. Highlight AAccttuuaattoorr mmaaiinntt.. and press EEnntteerr.
3. In actuator maintenance you will see the following options:
Both hilos up
Both hilos down
Hilo foot up
Hilo foot down
Hilo head up
Hilo head down
Foot up
Foot down
Head up
Head down
Knee up
Knee down
4. Use the up and down arrow buttons to highlight a movement and press and hold EEnntteerr to move the actuator.
SSccaallee mmaaiinntteennaannccee
Scale maintenance provides access to the following:
Gross weight
Net weight
Center of gravity
Load cells foot end
Load cells head end
Enable NAWI Scale
Calibrate scale
To access scale maintenance:
1. Enter the
Maintenance menu
(page 20).
2. Highlight SSccaallee mmaaiinntt.. and press EEnntteerr.
3. Use the up and down arrow buttons to highlight each number.
4. Press EEnntteerr to increase the number.
NNoottee -- The number will increase to 9 until it resets to 0.
5. The scale maintenance code is as follows: 4, 3, 2
5900-009-002 Rev A.0 21 EN

Other manuals for Stryker MV3 5900

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker MV3 5900 and is the answer not in the manual?

Stryker MV3 5900 Specifications

General IconGeneral
BrandStryker
ModelMV3 5900
CategoryMedical Equipment
LanguageEnglish

Related product manuals