SSuummmmaarryy ooff ssaaffeettyy pprreeccaauuttiioonnss
Always read and strictly follow the warnings and cautions listed on this page. Service only by qualified personnel.
WWAARRNNIINNGG
• Always use this product for bariatric use only per the details on the product labeling and specifications from this manual.
• Do not use this product for psychiatric, pediatric, or home health care use.
• Do not use this product in an oxygen rich environment.
• Always use Stryker approved mattresses that have been tested for compatibility with the product frame to avoid the risk
of patient entrapment.
• Do not lower the litter too low or the litter may not be supported.
• The use of accessories, transducers, and cables, other than those specified or provided by the manufacturer, could
result in increased electromagnetic emissions or decreased electromagnetic immunity and result in improper operation.
• Portable RF communications equipment, including peripherals such as antenna cables and external antennas, should
be no closer than 12 inches (30 cm) to any part of MMVV33, including cables specified by the manufacturer.
• Avoid stacking or placing equipment adjacent with other equipment to prevent improper operation of the products. If
such use is necessary, carefully observe stacked or adjacent equipment to make sure that they are operating properly.
CCAAUUTTIIOONN
• Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in this
manual.
• Do not modify the product or any components of the product. Modifying the product can cause unpredictable operation
resulting in injury to patient or operator. Modifying the product also voids its warranty.
• Always make sure that the product is clear of obstacles before you use motion functions.
• Always calibrate the product angles when you replace the main control board or any actuator or product damage may
occur.
• Always use ESD protective equipment before opening antistatic bags and servicing electronic parts.
• Do not place unprotected circuit boards on the floor.
• Always perform this procedure in a secure location with the floor jack secure to the base frame before you remove the
caster to prevent the risk of injury.
• Do not put your hands or feet below the caster when you remove or attach the caster to prevent the risk of injury.
• Always leave the bolts loose in the support holes, so the bolt head is out far enough to stop the lift mechanism slide
block.
EN 4 5900-009-002 Rev A.0