0702-014-700 Rev-B 2008/04
www.stryker.com 9
Specifications
Guidance and manufacturer's declaration - electromagnetic immunity
TheNeptune 2 Dockers,REF0702-014-000andREF0702-015-000,areintendedforuseintheelectromagneticenvironmentspecifiedbelow.Thecustomeror
the user of the Neptune 2 Dockershouldassurethatitisusedinsuchanenvironment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostaticdischarge
(ESD)
IEC61000-4-2
±6kVcontact
±8 kV air
±2,4,6kVcontact
±2, 4, 8 kV air
Floorsshouldbewood,concreteorceramictile.Iffloorsarecoveredwith
syntheticmaterial,therelativehumidityshouldbeatleast30%.
Electricalfasttransient/
burst
IEC61000-4-4
±2kVforpowersupply
lines
±1kV for input/output
lines
±2kVforpowersupply
lines
±1 kV for input/output
lines
Mainspowerqualityshouldbethatofatypicalcommercialorhospital
environment.
Surge
IEC61000-4-5
±1kVdifferentialmode
±2 kV common mode
±1kVdifferentialmode
±2 kV common mode
Mainspowerqualityshouldbethatofatypicalcommercialorhospital
environment.
Voltagedips,short
interruptionsandvoltage
variations on power
supplyinputlines
IEC61000-4-11
<5%U
T
(>95%dipinU
T
)
for0,5cycle
40%U
T
(60%dipinU
T
)
for5cycles
70%U
T
(30%dipinU
T
)
for25cycles
<5%U
T
(>95%dipinU
T
)
for5sec
95%Reduction(10ms)
60%Reduction(100ms)
30%Reduction(500ms)
95%Reduction(5s)
Mainspowerqualityshouldbethatofatypicalcommercialorhospital
environment.IftheuseroftheNeptune 2 Docker requires continued
operation during power mains interruptions, it is recommended that the
Neptune 2 Dockerbepoweredfromanuninterruptiblepowersupplyora
battery.
Powerfrequency
(50/60Hz)
magneticfield
IEC61000-4-8
3A/m 3A/m
@50Hz
CRT1A/m
Powerfrequencymagneticfieldsshouldbeatlevelscharacteristicsofa
typicallocationinatypicalcommercialorhospitalenvironment.
NOTE:U
T
isthea.c.mainsvoltagepriortoapplicationofthetestlevel.