8 www.stryker.com
2008/04 0702-014-700 Rev-B
Guidance and manufacturer's declaration - electromagnetic immunity
TheNeptune2Dockers,REF0702-014-000andREF0702-015-000,areintendedforuseintheelectromagneticenvironmentspecifiedbelow.Thecustomeror
the user of the Neptune 2 Docker shouldassurethatitisusedinsuchanenvironment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
ConductedRF
IEC61000-4-6
3Vrms
150kHzto80MHz
3Vrms
150kHzto80MHz
PortableandmobileRFcommunicationsequipmentshouldbeusedno
closertoanypartoftheNeptune2Docker,includingcables,thanthe
recommendedseparationdistancecalculatedfromtheequationapplicable
to the frequency of the transmitter.
Recommended separation distance
d=1.67√P
d=1.67√P
80MHzto800MHz
d=2.33√P
800MHzto2.5GHz
Where Pisthemaximumoutputpowerratingofthetransmitterinwatts
(W)accordingtothetransmittermanufacturerandd is the recommended
separationdistanceinmeters(m)
Interferencemayoccurinthevicinityofequipmentmarkedwiththe
followingsymbol:
RadiatedRF
IEC61000-4-3
3V/m
80MHzto2.5GHz
3V/m
80MHzto2.5GHz
NOTE1:At80MHzand800MHzthehigherfrequencyrangeapplies.
NOTE2:Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreflectionfromstructures,objectsand
people.
Specifications
Guidance and manufacturer's declaration - electromagnetic emissions
TheNeptune2Dockers,REF0702-014-000andREF0702-015-000,areintendedforuseintheelectromagneticenvironmentspecifiedbelow.Thecustomeror
the user of the Neptune 2 Docker shouldassurethatitisusedinsuchanenvironment.
Emissions test Compliance Electromagnetic environment - guidance
RFemissions
CISPR11
Group1 TheNeptune2Docker usesRFenergyonlyforitsinternalfunction.Therefore,itsRF
emissionsareverylowandarenotlikelytocauseanyinterferenceinnearbyelectronic
equipment.
RFemissions
CISPR11
ClassA TheNeptune2Docker issuitableforuseinallestablishmentsotherthandomestic
establishmentsandthosedirectlyconnectedtothepubliclow-voltagepowersupply
networkthatsuppliesbuildingsusedfordomesticpurposes.
Harmonicemissions
IEC61000-3-2
ClassA
Voltagefluctuations/flicker
emissions
IEC61000-3-3
Complies