Figure 32: Battery Power LED Indicator
A
CHECKING THE COT BATTERY POWER LEVEL
Power-LOAD automatically charges the Power-PRO SMRT Pak battery when the cot is locked into Power-LOAD in
the transport position (no cable or connectors required). The cot battery LED indicator momentarily flashes green to
signify that it is charging.
To check the battery power level, press the retract ()
button (2) on the cot control switch to activate the cot
battery LED indicator (A) as shown in Figure 32.
The cot battery LED indicator is located at the Power-
PRO foot end control enclosure (shown as a battery
symbol).
Ǩɣ The LED is solid green when the battery is fully
charged or has adequately charged battery
power.
Ǩɣ The LED flashes red when the battery needs to
be recharged or replaced.
Ǩɣ The LED is solid red to indicate a battery error.
See the SMRT Power System Operations/Maintenance
manual for additional SMRT Pak and SMRT Charger
information.
See the Power-PRO Operations/Maintenance manual
for additional cot battery information.
Notes:
Ǩɣ Automatic charging will only occur with SMRT Pak batteries.
Ǩɣ Only use Stryker-approved batteries with Power-LOAD.
Power-PRO Cot User Controls

Other manuals for Stryker Power-LOAD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Power-LOAD and is the answer not in the manual?

Stryker Power-LOAD Specifications

General IconGeneral
Load Capacity700 lbs (317.5 kg)
Battery TypeSealed Lead Acid
Charging Time8 hours
CompatibilityStryker ambulance cots
Power SourceBattery
Lifting MechanismHydraulic