27
Note If errors (F1-F8) persist, refer to the Service and Claims section of
this manual to obtain repair service.
Note If disturbance occurs on the video monitor or any other
electronic instrument, the user should:
Caution Ensure the Probe cable is not near any other instrument cables.
Caution Ensure the Endoscope has a non-metallic coupling ring (i.e.
Stryker CE marked Endoscopes).
NON-RECOVERABLE ERRORS & CORRESPONDING CODES
Error Code Reason for Error Solutions
F1 AC voltage incorrect
• Cycle power to the unit (turn the
unit off and then on)
F4 COP Watchdog • Same as above
F5 Oscillator fault • Same as above
F6 Software fault • Same as above
F7 Hardware fault • Same as above
F8 Power-on self test error • Same as above
SOLUTIONS
Problem Possible Solutions
System does not power uo after power switch is
pressed
• Check that power cord is properly
connected to Generator and into
an appropriately grounded outlet.
Green footswitch indicator light does not illuminate
• Check that Footswitch is properly
connected to the Generator and
no damage exisits to cable or
connector.
•Change Footswitch
Green Probe indicator light does not illuminate
• Check that Probe is properly
connected to the Generator and
no damage exisits to cable or
connector.
•Change Footswitch