27
Note If errors (F1-F8) persist, refer to the Service and Claims section of
this manual to obtain repair service.
Note If disturbance occurs on the video monitor or any other
electronic instrument, the user should:
Caution Ensure the Probe cable is not near any other instrument cables.
Caution Ensure the Endoscope has a non-metallic coupling ring (i.e.
Stryker CE marked Endoscopes).
NON-RECOVERABLE ERRORS & CORRESPONDING CODES
Error Code Reason for Error Solutions
F1 AC voltage incorrect
Cycle power to the unit (turn the
unit off and then on)
F4 COP Watchdog Same as above
F5 Oscillator fault Same as above
F6 Software fault Same as above
F7 Hardware fault Same as above
F8 Power-on self test error Same as above
SOLUTIONS
Problem Possible Solutions
System does not power uo after power switch is
pressed
Check that power cord is properly
connected to Generator and into
an appropriately grounded outlet.
Green footswitch indicator light does not illuminate
Check that Footswitch is properly
connected to the Generator and
no damage exisits to cable or
connector.
•Change Footswitch
Green Probe indicator light does not illuminate
Check that Probe is properly
connected to the Generator and
no damage exisits to cable or
connector.
•Change Footswitch