OPERATING PROCEDURES
24 OP CIC1200E Rev4.0 06-09 Chapter 4
4
4.7.4.3. UVPR
Meaning: Input signal of calibrated UV-probe con-
nected to socket.
UV-PROBE in the front panel will be displayed.
Function:
The alphanumeric display shows: Active modes re-
sp. LEDs in the annunciator group DISPLAY.
_ _ _ _ UVPR CHANNEL 2
CALIB
Press START
_ _ _ _ UVPR CHANNEL 1
CALIB
To leave this menu press CALIB
_ FINISH _
Press 8x CALIB:
4.7.4.4. CAL1 RES
Meaning: RESet CALibration data Channel 1
Press 8x CALIB
4.7.4.5. CAL2 RES
Meaning: RESes CALibration data channel 2
Function:
CAL1 RES _ CHANNEL 1
CALIB
Press START
CAL1 RES? CHANNEL 1
CALIB
Press START
DONE _ _ _ _
To leave this menu press CALIB.
_ STO RED _
Use the same procedure for CHannel 2 (CAL2RES).
4.7.4.6. Exit Calibration Menu
Press 10x CALIB
EXIT _ _ _ _
and press START.