EasyManua.ls Logo

SUSS CIC1200 - Calibration of the Built-In UV Intensity Meter

SUSS CIC1200
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OPERATING PROCEDURES
26 OP CIC1200E Rev4.0 06-09 Chapter 4
4
4.9. Calibration of the Built-In
UV Intensity Meter
Calibrate the built-in UV intensity meter using the
SUSS UV Intensity Meter (UV1000) and the UV
probe 405[nm] as an external reference.
Open the shutter of the lamp house.
Measure the actual UV intensity with the SUSS
UV Intensity Meter equipped with a correspond
-
ing probe head by placing it in the center of the
chuck. Remember the actual reading of the
meter - better: write it down.
Now connect probe head to the socket of the CIC
labeled UV-PROBE. Place it in the same position
as before. (better: Leave it after Step2) Activate
the calibration mode with key CALIB (press key
for 1-2 sec) Press CALIB until the display reads
UVPR. Press START. The alphanumeric display
shows UVP and the corresponding value of the
measured light intensity.
Check reading of the alphanumeric display (step
3) and compare it with the value for the light in
-
tensity obtained in step 2. If necessary adjust the
value for the LIGHT INTENSITY with the trim
-
ming potentiometer (see adjacent picture). Cali-
bration is accomplished when the numerical
value of the light intensity as given by the refer
-
ence meter (step2) is equal to the reading of the
alphanumeric display LIGHT INTENSITY of the
CIC1200
Control Panel of the CIC1200
The calibration point (potentiometer) for the built-in
UV Intensity Meter is indicated by „1“.
1