EasyManua.ls Logo

Sutter Instrument P-97 - Glass, Filaments, and Technical Support; System Descriptions: Front Panel and Mechanics

Sutter Instrument P-97
66 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2
1.2 Glass Capillary & Heating Filament Specifications
1.2.1 Glass Capillary
The P-97 micropipette puller is designed for use with aluminosilicate, borosilicate or other
lower melting-point glass tubing or rod ranging from 0.6 to 3.0 mm in diameter. This
instrument does not pull quartz glass. Examples of the specific types and sizes of glass that
can be used with the P-97 are listed in the Sutter Instrument Company catalogue that was
included with this instrument or can be viewed on Sutter Instrument’s web site at
www.sutter.com. Any glass with comparable technical specifications can be used with the P-97.
1.2.2 Heating Filament
The type and size of glass that you choose may require a Heating Filament other than the
one installed in your puller at the factory. Please refer to the Heating Filament section of this
manual to determine the appropriate style and size of filament necessary for pulling the
specific glass you would like to use. This instrument is designed to accommodate any of the
Sutter Instrument Trough-type or Box-type filaments that are shown in the Sutter
Instrument catalogue. This selection of replacement filaments can also be viewed on Sutter
Instrument’s web site at
www.sutter.com.
ONLY USE SUTTER-SUPPLIED REPLACEMENT HEATING FILAMENTS IN THIS
INSTRUMENT!
Systematic instructions for replacing the Heating Filament can be found in the Maintenance
section of this manual.
1.3 Technical Support
Unlimited technical support is provided by Sutter Instrument Company at no charge to our
customers. Our technical support staff is available between the hours of 8:00 AM and 5:00
PM (Pacific Standard Time) at (415) 883-0128. You may also Email your queries to
info@sutter.com
.
1.4 System Description – Front Panel
Figure 1-1. P-97 Front panel.
P-97 FLAMING/BROWN MICROPIPETTE PULLER OPERATION MANUAL – REV. 2.43 - DOM (20161118)