26
Line 2 indicates the measured steam temperature at the steam tank temperature probe.
Line 3 is the oset used to calibrate the steam tank temperature and pressure so that when the steam tank is
set at 250°F, there is about 1.3 bar pressure on the gauge.
Line 4 of this display screen is showing a 5 second delay. The ll probe will wait this long before turning the
steam tank ll valve on or o. Add me here if the machine is in an unstable installaon such as a food truck
or catering cart. There is no need to drop this me below 5 seconds.
Synesso keeps a log of the factory osets for each machine. Contact Synesso Technical Support if you need to
obtain the original factory-determined oset values.
Menu Level 2: Steam Oset
PROGRAMMING
Menu Level 2: Calibrate Flowmeters
The owmeter calibraon feature is made available with rmware v2.64.
The feature allows for technicians and advanced users to calibrate the owmeters in order to achieve ex-
treme accuracy in their output volumes or to recalibrate owmeters that have become less accurate over
me due to mineral buildup.
Line 3 allows the operator to begin the calibraon. Before performing the calibraon please read the in-
strucons on page 34. Instrucons should be carefully followed in order to ensure proper volumetric func-
oning of the machine.
To cycle to the next display screen, press the down buon.