maintenance/
troubleshooting
90
SYNRAD OEM v40 Operator’s Manual Version 4
Maintenance
3
Remove loose contaminants from the optic by holding a clean air bulb at an angle to the optic and
blow a stream of air at a glancing angle across the lens surface. Repeat as necessary.
4
Dampen a cotton ball or ued cotton swab with the selected cleaning agent. Ethyl alcohol (least
aggressive) is best for initial surface cleaning. Acetone (moderately aggressive) is best for oily residue
or minor baked-on vapors and debris.
Important Note: If acetone is used as a cleaning solvent, a second follow-up cleaning of the optical
surface using ethyl alcohol is required.
5
Gently wipe the optical surface with the damp cotton beginning in the center of the optic and work-
ing outward. Do not rub or apply any pressure, especially when using a cotton swab. Lightly drag
the cotton ball or swab without applying any downward pressure.
Note: Use a clean cotton ball or ued swab on each pass. The cotton will pick up and carry surface
contaminants that may scratch optical surfaces if reused.
6
Carefully examine the optic under a good light. Certain contaminants or damage such as pitting
cannot be removed. In these cases the optic must be replaced to prevent catastrophic failure.
7
Repeat Steps 4 through 6 as required, removing all traces of contaminants and deposits.