CHAPTER 1 INSTALLATION
13
1.2 Installation Conditions
Before installing the instrument, please confirm the following installation
conditions.
The requirements for UV2600 work power supply are as follows:
1)Voltage:AC100-240V, fluctuation within ±10% is allowed.
2)Frequency:50Hz or 60Hz,fluctuation within ±4% is allowed.
3)Power:more than 600VA is required. If other devices share the same power
supply,it needs more than 1KVA.
4)Grounding wire:Grounding resistance must be less than10Ωinstalled as per
relative standards.
The requirements for work table are as follows:
1)Work table for UV2600 spectrophotometer should be horizontal and reliable,
withstanding weight of more than 85kg, with a width of more than 800mm, a
depth of more than 750mm.
2)If other devices will be put on the work table as well, the area required
should be more.
3)Leave space of more than 200mm at both sides of the instrument. Avoid
instrument touching to the wall directly.
4)The height of UV2600 spectrophotometer is about 270mm. For easy
operation, select suitable height of work table.