CHAPTER 4 MAINTENANCE
34
4.3 Troubleshooting
4.3.1 Error Information
Symptom
Cause
Solution
ROM
ROM broken
Contact to serviceman
RAM
RAM broken
Contact to serviceman
6561nm
6561nm not found
Check whether sponge in
viewfinder has been taken out,
whether any material in sample
compartment, and do
wavelength calibration again.
Wavelength
drive
Wavelength drive error
Contact to serviceman
Baseline
Current baseline parameter not
complied with setting parameter
Reset parameters for baseline
calibration.
D2
When setting D2 lamp ON or need to
use D2 lamp, lamp is not ON.
Turn on D2 lamp
WI
When setting WI lamp ON or need to
use WI lamp, lamp is not ON.
Turn on WI lamp