CHAPTER 1 INSTALLATION
16
1.5 Instrument Installation
WARMING
The Voltage is AC100~240V, which may cause electric shock. Please
connect instrument power cord after instrument installation has been
finished.
To avoid shock during transportation, sponge has been used for fixing
viewfinder in UV2600. Before powering on the instrument, first open cover of
light source compartment and take out sponge in viewfinder. Otherwise, it may
have influence on instrument initialization check and normal use. The
procedures for taking out sponge are as follows
1Unscrew fix Screw on cover of light source compartment and remove cover
of light source compartment.
2Take out sponge inside viewfinder.
3Install cover and tighten the screw.
1.6 Check of Voltage and Fuse
WARMING
The Voltage of instrument is AC100-240V, which may cause electric
shock. Before selecting and confirming the voltage used, first confirm
power cord has not been connected to the instrument.
According to the voltage of work power supply, select and confirm the voltage
level allowed besides UV2600 plug of power supplyPOWER. Otherwise,
instrument failure or accident may occur.
Figure 1-3 Power supply of UV2600 Spectrophotometer
Fuse