UV2600 instrument has high precision. The purpose of maintenance is to
make spectrophotometer work under good status.
Sample compartment is easy to be contaminated. If sample is spilled inside the
compartment, make sure to wash the compartment. The wash procedures are
as follows. Reassemble the sample compartment after it is dry. (Figure 1-4 to
Figure 4-5).
(1) Open cover of sample compartment.
(2) Unscrew the screw of cuvette holder (one screw).
(3) Hold front board of sample compartment, and take out cuvette holder
(Note:there is assigned bolt).
(4) Clean and dry cuvette holder and inner of sample compartment.
(5) Install cuvette holder back to sample compartment and tighten the screw.