TECNA WELD TESTER TE1700
Signals that the progressive counter used to save the
.CSV (Comma-Separated Values) file to pen-
drive was
reset.
Signals that the internal memory in which the spots to
be used for statistics are stored was cleared.
The instrument is too near to the secondary area of the
spot welder and the magnetic field generated interferes
with the correct execution of the measurement. Move
the instrument away from the secondary circuit of the
spot welder.
Signals that the acquisition time of the measurement
process just performed was too long. It should be
reduced.
Signals that the execution speed of the current welding
spots is too high. Reduce the speed.
Signals that the instrument is running a firmware
upgrade and needs to be switched off and back on.