Performance Check—2465B/2467B Service 
HORIZONTAL 
Equipment Required (see Table 4-1) 
Primary Leveled Sine-Wave Generator (Item 2) 
Calibration Generator (Item 3) 
Time-Mark Generator (Item 6) 
T-Connector (Item 8) 
Precision 50
 <2
 BNC Cable (2 required) (Item 10) 
Dual Input Coupler (Item 11) 
Pulse Generator (Item 24) 
Initial Control Settings. 
Control settings not listed do not affect the procedure. 
Set: 
NOTE 
Select channels to set VOLTS/DIV. 
VOLTS/DIV 
CH 1 and CH 2 
CH 1 VAR 
CH 3 and CH 4 
0.5 V 
In detent 
0.1 V 
VERTICAL MODE 
CH 1 
CH 2, CH 3, CH 4, 
ADD,
 and INVERT 
CHOP/ALT 
20 MHz BW LIMIT 
Input Coupling 
CH 1 and CH 2 
Horizontal 
A SEC/DIV 
SEC/DIV VAR 
X10MAG 
TRACE SEP 
Delta 
AV and At 
TRACKING 
On 
Off 
ALT 
Off 
50S2DC 
200 ns (knob in) 
In detent 
Off 
Fully CW 
Off (press and release 
until associated readout 
is off) 
Off 
Trigger 
HOLDOFF 
LEVEL 
SLOPE 
MODE 
SOURCE 
COUPLING 
B ENDS A 
Midrange 
+ (plus) 
AUTO LVL 
VERT 
DC 
1.
 Check Horizontal Display Modes (A, A INTEN, 
ALT, and B). 
a. Use a 50
 12
 BNC cable to connect 200 ns time mark-
ers from the Time-Mark Generator to the CH 1 OR X input 
connector. 
b. Adjust the TRIGGER LEVEL control as necessary for 
a stable signal display. 
c. Pull the SEC/DIV knob out and set the B TRIGGER 
MODE to RUN AFT DLY. 
NOTE 
On CTT instruments, rotate the A REF OR DLY 
POS control for the specified delay. As the control is 
rotated, the readout delay value will be followed by 
the word
 "SET".
 This value shows the approximate 
delay. A few seconds after control movement has 
stopped, the word "SET" will disappear and the 
readout delay value as measured by the CTT will 
appear. This is normal operation and not cause for 
concern. 
d.
 Set the A REF OR DLY POS control for a DLY 
readout of approximately 1000 ns. 
e. VERIFY—An intensified zone appears on the 
displayed signal near graticule center. The INTENSITY 
control may need adjustment. 