Specification—2465B/2467B Service
Table 1-1 (cont)
Characteristics
X10 MAG Accuracy
500 ms or 200 ms/div Timing Accuracy (A Sweep
only)
SEC/DIV VAR Timing Accuracy
Timing Accuracy (-15°C to +15°C and +35°C to
+ 55°C)
At Readout Resolution
At Range
Sweep Delay Range
Delay Jitter
2467B
2465B
Horizontal POSITION Range
X-Y Operation
X-Axis Deflection Factor Range, Variable, and
Input Characteristics
Deflection Factor Accuracy
X-Axis Bandwidth
Phase Difference Between X and Y with BW
Limit Off
X-Axis Low-frequency Linearity
Performance Requirements
Add ±0.5% of time interval to unmagnified Sweep and At Cursors
specifications. Exclude the first 0.5 division after the sweep starts
(the first 0.5% of the full 100 division sweep).
Add ± 0.5% of interval to specifications for A SEC/DIV at 100 ms
or faster.
Add 2% of time interval to sweep accuracy specifications when
VAR is out of detent.
Add ±0.2% of time interval to all At and delay specifications. Add
±0.5%
of interval to sweep accuracy specification.
3
Greater of either 10 ps or 0.025% of full scale.
3
±10 times A-SEC/DIV setting with Cursors, ±9.95 times A-
SEC/DIV setting with Sweep Delay.
3
0 to 9.95 times the A SEC/DIV setting, from 500 ms to 10 ns. A-
Sweep triggering event is observable on B Sweep with zero delay
setting for A SEC/DIV settings 10
MS
or faster.
3
Within
0.01%
(one part or less in 10,000) of the maximum available
delay, plus 100 ps.
a
Within 0.004% (one part or less in 25,000) of the maximum
available delay, plus 50 ps.
3
Start of
1
ms per division sweep can be positioned from right of
graticule center to at least 10 divisions left of graticule center.
Some portion of
1
ms per division sweep is always visible with X10
MAG
off.
3
Same as Channel 1.
Same as Channel 1.
3
Dc to 3 MHz.
<1 ° from dc to 1 MHz; «s3° from 1 MHz to 2 MHz.
0.1 division or less compression or expansion of a two-division,
center-screen signal when positioned within the graticule area.
"Performance requirement not checked in manual.
1-8