Turn Off Time. Turn Off Time is the time taken to get the output voltage to low level (close to zero) after the input voltage is
removed.
â– 
The Maximum Voltage value should be greater than the Input Trigger level.
â– 
The Maximum Voltage level value sets the waveform vertical settings to approximately one-fourth of the specified level.
â– 
The Maximum Time field sets the horizontal units so as to display the entire time in which the output is supposed to reach
zero.
â– 
The Input Trigger level is set based on the 50 % level of the input channel.
â– 
Turn Off Time sets the trigger slope to rising edge.
Measurement algorithms