Compliance Info
rmation
Compliance In
formation
This section lists the EMC (electromagnetic compliance), s afety, and environmental standards with which the instrument
complies.
EMC Compliance
EC Declarati
on of Conformity – EMC
Meets intent of Directive 2004/108/EC for Electromagnetic Compatibility. Compliance was demonstrated to the following
specifications as listed in the Official Journal of the European Communities:
EN 61326-1:2006, EN 61326-2-1:2006. EMC requirements for electrical equipment for measurement, control, and
laborator
yuse.
1
,
2
,
3
CISPR 11:2
003. Radiated and conducted emissions, Group 1, Class A
IEC 61000
-4-2:2001. Electrostatic discharge immunity
IEC 61000
-4-3:2002. RF electromagnetic field immunity
4
IEC 61000
-4-4:2004. Electrical fast transient/burst immunity
IEC 6100
0-4-5:2001. Power line surge immunity
IEC 6100
0-4-6:2003. Conducted RF immunity
4
IEC 6100
0-4-11:2004. Voltage dips and interruptions immunity
5
EN 61000-3-2:2006. AC power line harmonic emissions
EN 6100
0-3-3:1995.
Voltage changes, fluctuations, and flicker
European Contact.
Tektronix UK, Ltd.
Western Peninsula
Western Road
Bracknell, RG12 1RF
United Kingdom
1
This
product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic interference.
2
Emissions which exceed the levels required by this standard may occur when this equipment is connected to a test object.
3
If interconnect cables are used, they must be low–EMI shielded cables such as the following Tektronix part numbers or their
equivalents: 012-0991-01, 012-0991-02 or 012-0991-03 GPIB Cable; 012-1213-00 (or CA part number 0294-9) RS-232 Cable;
012-
1214–00 Centronics Cable; or LCOM part number CTL3VGAMM-5 VGA Cable. Use an 012-0482-00 cable for the Ref Out connector.
4
The performance criterion for when the oscilloscope is subjected to the continuously present electromagnetic phenomenon:
10 mV
/division to 1 V/divis ion: ≤0.4 division waveform displacement or ≤0.8 division increase in peak–to–peak noise Performance
criterion for when the oscilloscope is subjected to transient electromagnetic phenomenon: Temporary, self–recoverable degradation
or loss of performance is allowed, but no change of actual operating state or loss of stored data is a llowed.
5
Performance Criterion C applied at the 70%/25 cycle Voltage-Dip and the 0%/250 cycle Voltage-Interruption test levels
(IEC 61000-4-11). If the instrument powers down upon a voltage dip or interruption it will take longer than ten seconds to reboot.
DPO7000 S e ries and DPO/DSA70000/B Series Quick Start User Manual vii