Behavior. Use the Window trigger to trigger the instrument when the input signal rises above an upper threshold level or falls
below a lower threshold level. After setting these levels, you can specify whether you want to trigger the instrument as the signal
is entering or leaving the threshold window. The trigger can also be qualified by the logical state of other channels or by a bus
pattern (MSO70000C/DX Series only). When you click on the trigger window controls, the trigger graphic shows the trigger
sequence.
You can further qualify the trigger event in terms of time by using the Inside > t or Outside > t selections, or by the logical state of
other channels using the Trigger When Logic option.
The Trigger When controls change as you select different combinations of options.
What do you want to do next?
Learn more about other trigger types.
Return to the Trigger Setup control window.
Trigger graphic
The trigger graphic displays a visual indicator of the trigger criteria. Depending on the trigger type, the instrument will update the
graphic as you define the trigger parameters.
Sequence triggering
Set up sequential triggering (Horizontal delay off)
From the Trig menu, select A - B Trigger Sequence; then open the A->B Seq tab.
Trigger setups
DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series 441

Table of Contents

Other manuals for Tektronix DPO7054C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO7054C and is the answer not in the manual?

Tektronix DPO7054C Specifications

General IconGeneral
BrandTektronix
ModelDPO7054C
CategoryTest Equipment
LanguageEnglish

Related product manuals