Parallel.For(0, vCount, vv =>
{
output[vv, hh] = output[vv, hh + 1];
});
}
If the output type of a plugin is IWaveformDB, when the plugin is called the output waveform is populated with the values the
plugin returned the last time it was called. If this is the first time the plugin was called, all of the values inside the IWaveformDB
will be zero (displays as clear).
The advanced user can change the vertical scale and position, as well as the horizontal scale and spacing by setting
Vertical.Scale, Vertical.Position, Horizontal.Scale, and Horizontal.Spacing in the IWaveformDB.
IString. A plugin may use an unlimited number of IStrings as input. An IString is simply a string and is used to pass meta data or
additional information into a plugin.
Oscilloscope reference
710 DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series