EasyManua.ls Logo

Tektronix MSO4104

Tektronix MSO4104
59 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Verification
Table 12: Maximum Bandwidth Frequency worksheet (cont.)
Model: DPO4054, MSO4054
1MΩ 2 mV/div
300 MHz
1MΩ 1 mV/div
175 MHz
Model: DPO4034, DPO4032, MSO4034, MSO4032
50 Ω 5 mV/div
350 MHz
50 Ω 2 mV/div
350 MHz
50 Ω 1 mV/div
200 MHz
1MΩ 5 mV/div
350 MHz
1MΩ 2 mV/div
300 MHz
1MΩ 1 mV/div
175 MHz
1
For DPO4104, MSO4104, DPO4054, and MSO4054 bandwidth verification, use 380 MHz, rather than 500 MHz,
on the 5 mV/div vertical scale due to an impedance mismatch between the calibrator and the oscilloscope.
When the calibrator is set to 1 MΩ load, it has a Thevenin equivalent 25 Ω source impedance. Passing the test
with a 380 MHz signal verifies 500 MHz performance with a P6139A probe on models DPO4104, MSO4104,
DPO405
4, and MSO4054.
16. Use the values of V
bw-pp
and V
in-pp
obtained above and stored in the test record
to cal
culate the Gain at bandwidth with the following equation:
Gain = V
bw-pp
/ V
in-pp
To pass the performance measurement test, Gain should be ≄ 0.707. Enter
Gain in the test record.
17. Repeat steps 9 through 16 for the other oscilloscope volts/div settings listed
in the test record.
18. Set the calibrator to 1 MΩ output impedance to generate a sine wave.
19. Push the channel button (1,2,3,4) for the same channel that you used in step 3.
20. Change the oscilloscope impedance to 1 MΩ.PushtheImpedance
lower-bezel button to select 1 M Ω.
21. Repeat steps 9 through 17.
22. Repeat steps 3 through 21 for each channel combination listed in the test
record and relevant to your model of oscilloscope (for example, 1,2,3,4).
40 MSO4000 and DPO4000 Series Specifications and Performance Verification

Other manuals for Tektronix MSO4104

Related product manuals