EasyManua.ls Logo

Tektronix RSA306 - Amplitude

Tektronix RSA306
69 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specications
Amplitude
RF input impeda
nce
50
RF input VSWR, typical 1.8:1 (10 MHz to 6200 MHz, reference level +10 dBm)
(Equivalent Return Loss: 11 dB)
Maximum RF input level without
damage
The maximum voltage or power that the RF input can withstand without creating
a shock hazard or damaging the input.
DC voltage
±40 V
DC
Ref Level –10 dBm +23 dBm (continuous or peak)
Ref Level < –10 dBm +15 dBm (continuous or peak)
Maximum RF input operating
level
The maximum level at the RF input for which the instrument will meet its
measurem
ent specications.
Center fr
equency < 22 MHz
+15 dBm
Center f
requency 22 MHz
+20 dBm
Amplitude accuracy at all center
freque
ncies
Refere
nce level +20 dBm to –30 dBm, alignment run prior to testing
Applies to corrected IQ data, with signal to noise ratios > 40 dB
Accuracy may degrade up to ±0.6 dB after storage at maximum storage temperature,
recove
rs within 24 hours
Cente
r frequency 9 kHz to
<3GHz
±2.0 d
B (18 °C to 28 °C)
±1.25 dB (18 °C to 28 °C), typical (95% condence)
±3.0 dB (–10 °C to 55 °C), typical
Center frequency 3GHzto
6.2 G
Hz
±2.75 dB (18 °C to 28 °C)
±2.0
dB (18 °C to 28 °C), typical (95% condence)
±3.0 dB (–10 °C to 55 °C), typical
Channel amplitude atness
Reference level +20 dBm to –30 dBm, alignment run prior to testing
Applies to corrected IQ data, with signal to noise ratios > 40 dB
±1.
0 dB (18 °C to 28 °C)
±2.0 dB, (–10 Cto55C), typical
±3.0 dB, 22 MHz - 24 MHz, (-10 Cto55C), typical
3 RSA306 Specications and Performance Verication

Other manuals for Tektronix RSA306

Related product manuals