Specications
Regulatory
Safety UL61010-1, CAN/CSA-22.2 No.61010-1, E N61010-1, IEC61010-1
Regional certications
Europe: EN613
26
Australia/New Zealand: AS/NZS 2064
EMC emissions
EN61000-3-2,
EN61000-3-3, EN61326-2-1
EMC imm unity EN61326–1/2, IEC61000-4-2/3/4/5/6/8/11
Environmen
tal performance
Temperature
Operating –10 °C to +55 °C (+14 °F to +131 °F)
Nonoperating
–51 °C to +71 °C (–60 °F to +160 °F)
Humidity, operating
5% to 95 ±5% RH (relative humidity) in the temperature range of +10 °C to 30 ° C
(+50 °F t
o8F)
5% to 75% ±5% RH from +30 °C to +40 °C (+86 °F to 104 °F)
5% to 45% RH above +40 °C to +55 °C (+86 °F to +131 °F)
<10 ° C hu
midity is uncontrolled; non-condensing
Altitude
Operating 9144 meters (30,000 feet)
Nonoperating
15,240 meters (50,000 feet)
Dynamics
Rand
om vibration:
nonoperating
0.030 g
2
/Hz, 10 Hz 500 Hz, 30 minutes per axis, three axes (90 minutes total)
Mech
anical shock: operating
Half-sine mechanical shocks, 30 g peak am plitude, 11 ms duration, three drops
in each direction of each axis (18 total)
Handling and transit
Ben
ch handling, operating
Per MIL-PRF-28800F Class 2 operating: Rotational-edge-drops of appropriate edges
on appropriate sides of the equipment
Transit drop, nonoperating
Per MIL-PRF-28800F Class 2 nonoperating: Transit drops onto six faces and four
co
rners of the equipment, from a height of 30 cm (11.8 in.) for a total of 10 impacts
9 RSA306 Specications and Performance Verication