Performance ver
ication
Phase noise
The intent of the Phase Noise test is to measure the phase noise level of the instrument. The phase noise specication does
not cover residual spurs. If the specic m easurement frequency results in measuring a r esidual spur that is visible above the
noise level, the phase noise specication applies not to the spur but to the noise level on either side of the spur. P lease
refer to the Spurious Response specications. Also, refer to the Spurious Response section of this procedure to determine
whether or not a residual spur is within the specication.
Connect the signal generator and RSA306 as shown in the following gure.
1. Reset the RSA306 to factory defaults (Presets > Main).
2. Run the RSA306 alignment procedure (Tools > Alignments > Align Now). Note: the Center frequency should be 1 GHz.
3. Set the signal generator CW frequency = 1GHz.
4. Set the signal generator CW amplitude = 0dBmat the RS306 input.
5. Select External Reference (Setup > Acquire > Frequency Reference > External)
6. Set trace detection = +PEAK (Setup > Settings > Traces > Detection)
7. Measure the CW amplitude for the following settings:
a. Trace Function = Avg (Vrms), 10 averages (Setup > Settings > Traces > Function: Avg (Vrms), count = 10)
b. Span = 100 kHz
c. RBW = 1kHz
d. Move MR marker to highest amplitude signal after 10 averages, write the marker value as the CW amplitude (for the
1kHzlter) in the measurement table. (See Table 14.)
8. Measure the CW amplitude for the following settings:
a. Span = 10 kHz
b. RBW = 100 Hz
c. Move MR marker to highest amplitude signal after 10 averages, write marker value as the CW amplitude for the
100 Hz lter and 10 Hz lter in the measurement table. (See Table 14.)
9. Set Trace detection = Avg (Vrms).
10. Set Marker function = Power (dBm/Hz) (Mar kers > Dene Markers > Readouts (near bottom of window) > P ower).
11. Turn on marker (Markers > De ne Markers > Add)
31 RSA306 Speci cations and Performance Verication

Other manuals for Tektronix RSA306

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix RSA306 and is the answer not in the manual?

Tektronix RSA306 Specifications

General IconGeneral
BrandTektronix
ModelRSA306
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals