Performance Verification
DANL - Preamp ON, LF
Path (Option 50/51 Only)
1. Reset the RSA51
00B to factory defaults: select Setup > Preset (Main).
2. Select Tools > Alignments and select Align Now.
3. Set the RSA5100B:
Reference Level
Setup > A mplitude > Internal Settings > Ref Level
–50 dBm
Internal Attenuator
Setup > Amplitude > Internal Settings > Internal
Attenuator
0 dB (Auto unchecked)
RF & IF Opti
mization
Setup > Amplitude > RF & IF Optimization
Minimize Noise
Internal
Preamp
Setup > Amplitude > Internal Settings > Internal
Preamp
ON (Internal Preamp box checked)
Center Frequency
Setup > S
ettings > Freq & Span > Center
1.1 MHz
Span
Setup > Settings > Freq & Span > span
100 kHz
RBW
Setup > Settings > BW > RBW
Auto (box checked)
Detec
tion
Setup > Settings > Traces > Detection
Avg (of logs)
Funct
ion
Setup > Settings > Traces > Function
Avg (of logs)
Count
Setup > Settings > Traces > Function
100 (Count box checked)
LF Path
Setup > Acquire > Input Params
Use Low Freq... (box checked)
4. Set the markers for Noise Mode (dBm/Hz) operation:
a. Select Markers located at the bottom right of the screen on the status bar
to display the Markers control bar.
b. Click the down arrow on the left side of the Markers control bar to view
the Select a Marker dropdown menu. Select Add Marker to add the
M
Rmarker.
c. Select Add Marker againtoaddtheM1marker.
RSA5100B Series Technical Reference 117