Operation
7. Click any point in the mask to select it.
The frequency and amplitude of the
selected poin
t are highlighted in the
graph.
8. Edit the point by doing one of the
following:
Drag and drop the point with the
mouse.
Enter the fr
equency and amplitude in
the table.
Adjust amplitude with the up/down
arrow keys and frequency with the
left/righ
tarrowkeys.
Quick Tips
Click Au to draw to generate a mask automatically adjusted to the Reference.
Right click on the display s creen to add a new point.
Click Add Points and then click on the graph to add a new point in the graph.
Click Next to select the next point in the graph.
Click the X (delete button) to delete the selected mask point.
Click Save to save the mask le and continue working with the mask editor.
Click Open to load and edit a previously saved mask le.
Dening Trigger Conditions
After you have set up the mask, you can select whether to trigger on the presence or absence of mask violations.
1. Select a condition.
The c
hoices are described in the
following table. (See Table 7.) The icons
at the bottom of the control panel identify
wha
t is considered true or false.
2. Cli
ck Triggered.
The instrument will trigger when the
trigger conditions are met.
52 RSA6100A Series & RS A5100A Series Quick Start User Manual