Operation
8. Increase the span in the Spectrum
display to 200 MHz.
Because Spect
rum is the selected
display, the acquisition parameters
adjust automatically to meet the needs
of the displa
y. The 200 MHz span
exceeds the real-time bandwidth of the
instrument. As a result, the instrument
will change t
o swept mode.
9. Look at the S
tatus bar. The readout
“Swept” indicates when acquisitions are
swept.
When the ac
quisitions are in “Real Time,”
all measurements usually run. If the
acquisitions are swept, only the selected
display ca
nrun.
10. Click the
Amplitude vs Time display to
make it the active display.
The acquisition switches back to
real-ti
me because the selected display
requires real-time. The data is now
suitable for the Amplitude vs Time
displa
y but doesn’t contain the full range
of frequencies needed by the Spectrum
display.
In this
case, the Spectrum is unable to
run.