Maintenance
TDS 200 Series Digital Oscilloscope Service Manual
639
Begin with this troubleshooting tree to isolate a problem to a specific trouble-
shooting procedure.
Check diagnostic screen
displayed at power-on.
Any
failed
messages?
Does
backlight
come on?
Is
the display
functioning
properly?
Is display
color abnormal
(dark blue, green, or
orange)?
Go to Troubleshooting
Display (page 642).
Does the
instrument
turn on?
Go to
Troubleshooting
Power Supply
(page 640).
Go to Checking
Probe Comp
(page 640).
Is there
a display?
Plug the instrument into a
proper power source and
press the power button.
Go to Troubleshooting
Backlight (page 641).
Yes
No
Yes
Yes
No
Yes
No
Yes
No
Yes
No
Go to Checking
Probe Comp
(page 640).
No
Instrument should be functioning.
Try the Other Troubleshooting
procedures (page 646) and the
Front-Panel Troubleshooting
(page 643) if other problems exist.
Begin Here