Probe Cal
TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–87
If the Re-use Probe Calibration data? menu (see Figure 3–45) is displayed,
choose one of the following options:
To use the Probe Cal data last stored for that channel, press OK Use
Existing Data
(side).
To erase the Probe Cal data last stored and use the probe uncompensated,
press OK Erase Probe Cal Data
(side).
To retain the Probe Cal data last stored and use the probe uncompensated,
press CLEAR MENU.
NOTE.
If the Re-use Probe Calibration data menu is displayed, do not select OK Use
Existing Data unless the calibration data was created using the currently installed
probe.
Figure 3–45: Re-use Probe Calibration Data Menu