EasyManua.ls Logo

Tektronix TDS 420A - Page 104

Tektronix TDS 420A
433 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Descriptions
2–90
TDS Family Oscilloscope Programmer Manual
Cursor
CURSor:FUNCtion
CURSor:PAIred:VDELTA?
CURSor
:
PAIred :
?
VDELTA
CURSOR:PAIRED:VDELTA?
might return 1.064E+00, indicating that the time between the paired cursors is
1.064 seconds.
CURSor:VBArs
Positions the vertical bar cursors and the CURSor:VBArs? query returns the
current vertical bar cursor settings for horizontal position, delta, cursor selection,
and units.
The position can appear in units of seconds, 1/seconds (hertz), or video line
numbers (with option 05). On the TDS 400A, position can also appear in units of
clocks or 1/clocks.
In extended-acquisition-length mode, the cursors are limited to the waveform
record section of the acquisition data (see Figure 2–4 on page 2–41).
Cursor
DATa:STARt, DATa:STOP, MEASUrement:GATing
CURSor:VBArs SNAp
CURSor:VBArs?
CURSor
:
VBArs
<Space>
?
SNAp
SNAp positions the vertical bar cursors at DATa:STARt and DATa:STOP.
Group
Related Commands
Syntax
Examples
Group
Related Commands
Syntax
Arguments

Table of Contents