Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–101
Table 2–30: DATa and WFMPre Parameter Settings
ATa N g e i g
WFMPre Settings
D
ATa
:E
N
Cd
g
S
e
tt
i
n
g
:ENCdg :BN_Fmt :BYT_Or
ASCIi ASC
N/A N/A
RIBinary BIN RI MSB
RPBinary BIN RP MSB
SRIbinary BIN RI LSB
SRIbinary BIN RP LSB
DATA:ENCDG RPBINARY
sets the data encoding format to be positive integer where the most significant
byte is transferred first.
DATA:ENCDG?
might return SRPBINARY for the format of the waveform data.
DATa:SOUrce
Sets or queries the location of the waveform data that is transferred from the
instrument by the CURVe? query. The source data is always transferred in a
predefined order regardless of the order they are specified using this command.
The predefined order is CH1 through CH4, MATH1 through MATH3, and then
REF1 through REF4.
In extended-acquisition-length mode, the MATH1 through MATH3 sources can
not be used. Also, in this mode, CH1 through CH4 refer to extended-acquisition-
length data. In contrast, when not in extended-acquisition-length mode, CH1
through CH4 refer only to the waveform record. For more details, see page 2–41.
Waveform
DATa:SOUrce <wfm>[<Comma><wfm>]...
DATa:SOUrce?
Examples
Group
Syntax