Command Descriptions
2–92
TDS Family Oscilloscope Programmer Manual
In extended-acquisition-length mode, the cursors are limited to the waveform
record section of the acquisition data (see Figure 2–4 on page 2–41).
Cursor
CURSor:VBArs:UNIts
CURSor:VBArs:POSITION<x> <NR3>
CURSor:VBArs:POSITION<x>?
CURSor
:
VBArs :
POSITION
<x>
?
<Space> <NR3>
<NR3> specifies the cursor position in the units specified by the CUR-
Sor:VBArs:UNIts command. The position is relative to the trigger position.
CURSOR:VBARS:POSITION2 9.00E-6
positions one of the vertical bar cursors at 9 ms.
CURSOR:VBARS:POSITION1?
might return 1.00E-6, indicating a vertical bar cursors is at 1 ms.
CURSor:VBArs:POSITION<x>Pcnt
TDS 400A Only
Sets or queries the position of the vertical bar cursors (x is either 1 or 2) in units
of % of record length.
Cursor
CURSor:VBArs:POSITION<x>Pcnt <NR3>
CURSor:VBArs:POSITION<x>Pcnt?