EasyManua.ls Logo

Tektronix TDS 420A - Page 123

Tektronix TDS 420A
433 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–109

might return the string  , showing that the DESER contains the binary
value 10111010.
DIAg:RESUlt:FLAg? (Query Only)
Returns the pass/fail status from the last diagnostic test sequence execution. The
DIAg:RESUlt:LOG? query can be used to determine which test(s) has failed.
Calibration and Diagnostic
DIAg:RESUlt:LOG?




 indicating that all of the selected diagnostic tests have passed.
 indicating that at least one of the selected diagnostic tests has failed.

returns either  or .
DIAg:RESUlt:LOG? (Query Only)
Returns the internal results log from the last diagnostic test sequence execution.
The list contains all modules and module interfaces that were tested along with
the pass/fail status of each.
Calibration and Diagnostic
DIAg:RESUlt:FLAg?

Group
Related Commands
Syntax
Returns
Examples
Group
Related Commands
Syntax

Table of Contents

Other manuals for Tektronix TDS 420A

Related product manuals