Command Descriptions
2–154
TDS Family Oscilloscope Programmer Manual
<Space>
<NR1>
HEADer
ON
OFF
?
ON or <NR1> 0 sets the Response Header Enable State to true. This causes the
digitizing oscilloscope to include headers on applicable query responses. You can
then use the query response as a command.
OFF or <NR1> = 0 sets the Response Header Enable State to false. This causes the
digitizing oscilloscope to omit headers on query responses, so that only the
argument is returned.
HEADER OFF
causes the digitizing oscilloscope to omit headers from query responses.
HEADER?
might return the value 1, showing that the Response Header Enable State is true.
HIStogram? (Query Only)
TDS 500C & 700C Only
Returns histogram parameters. In other words, it queries the state of all settable
histogram parameters.
Histogram
HIStogram?
HIStogram
?
HISTOGRAM?
might return
:HISTOGRAM:SOURCE CH1;MODE VERTICAL;DISPLAY LOG;BOXPCNT