Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–231
MEASUrement:METHod
Sets or queries the method used to calculate the 0% and 100% reference level.
This is equivalent to setting the High-Low Setup in the Measure menu.
Measurement
MEASUrement:METHod { HIStogram | MINMax }
MEASUrement:METHod?
MEASUrement :
METHod
<Space>
?
MINMax
HIStogram
HIStogram sets the high and low waveform levels statistically using a histogram
algorithm.
MINMax sets the high and low waveform levels to MAX and MIN, respectively.
MEASUREMENT:METHOD HISTOGRAM
specifies that the high and low reference levels are set statistically.
MEASUREMENT:METHOD?
returns MINMAX when the reference levels are set to MIN and MAX.
MEASUrement:REFLevel? (Query Only)
Returns the reference levels.
Measurement
MEASUrement:REFLevel?
MEASUrement :
REFLevel
?
Group
Syntax
Arguments
Examples