Command Descriptions
2–258
TDS Family Oscilloscope Programmer Manual
DELEte:WAVEform, SAVE:WAVEform:FILEFormat
SAVe:WAVEform <wfm>,{ REF<x> | <file path> }
<file path>
<wfm>
SAVe :
WAVEform
<Space>
REF
<x>
,
<wfm> is CH<x>, MATH<x>, or REF<x>, and is the waveform that will be
saved.
REF<x> is the location where the waveform will be stored.
<file path> (on instruments with the Options 1F or HD File System) is a
quoted string that defines the file name and path. Input the file path using the
form <drive>/<dir>/<filename>. <drive> and one or more <dir>s are
optional. If you do not specify them, the digitizing oscilloscope will write the
file to the current directory. <filename> stands for a filename of up to 8
characters followed by a period (“.”) and the proper 3-character extension.
Internal format waveforms use the “WFM” extension. The TDS 400A, 500C,
600B, and TDS 700C can also use a “CSV” extension for spreadsheet format files
or a “DAT” extension for MathCad format files.
The current directory refers to the name of a directory as returned by the
FILESystem:CWD query.
SAVE:WAVEFORM MATH2,REF1
saves the math 2 waveform in reference memory location 2.
SAVE:WAVEFORM MATH1,TEK00000.WFM
saves the math1 waveform to the file TEK00000.WFM in the default directory and
on the default drive.
SAVe:WAVEform:FILEFormat
Not on TDS 510A
Specifies the file format for saved waveforms.
Save and Recall
SAVE:WAVEform
Related Commands